HATP-4000 is a measuring instrument that automates the ultraviolet / visible spectroscopy (thermal oxidation method) of the water pollution test method. In the pretreatment step, potassium persulfate is added to the sample and heated at 120°C for 30minutes to oxidize and decompose the phosphorus compound into phosphate form. Oxidized phosphorus phosphate is colored with molybdenum blue and absorbance is measured. The absorbance of total phosphorus is measured at 880nm which is the maximum absorption wavelength of phosphorus phosphorus. HATP-4000 is KORBI’s new product which not only improved efficiency by adopting patented technology and our own developed oxidation tank, but also has sophisticated design and user’s convenience.
Characteristics
HATP-4000
Adoption of oxidation reaction chamber with improved efficiency
Application of multi-channel valve for quantitative distribution of samples and reagents
Built-in user-oriented operating program with enhanced security function
Built-in functions to enhance user convenience
Specification
-
- Measurement item
Total Phosphorus (TP)
-
- Measurement method
Absorbance photometry (Ascorbic acid reduction method / molybdic acid blue method)
-
- Measurement range
0~5mg/L (Adjustable)
-
- Wavelength
880nm
-
- Measurement interval
60 minutes (Adjustable)
-
- Detection limit
0.02mg/L F.S. or less
-
- Accuracy
±2% F.S.
-
- Reproducibility
±2% F.S.
-
- Linearity
±3% F.S.
-
- Zero/Span drift
±2% F.S.
-
- Oxidation method
120°C persulfate thermal oxidation
-
- Calibration
Auto/manual calibration by standard solution
-
- Sample/Reagent measuring method
Step motor pulse control
-
- Sample/Reagent distribution method
7-port distribution control method (Fixed quantity injection with multi-channel valve system)
-
- Operation system
Windows 7 Pro
-
- Output
DC 4~20mA, RS-232C
-
- Status display
Normal operation, on calibration, error
-
- Display
Color LED 10.4″ (Touch screen)
-
- Power
AC 220V, 50/60Hz
-
- Power consumption
Average 150W (Maximum 450W)
-
- Voltage regulation
3% or less
-
- Data storage capacity
HDD 100G or more
-
- Data output
USB terminal
-
- Enclosure/Size
Steel, 1.5T (Except paint) / 645(W) x 439(D) x 1600(H) mm
-
- Operating humidity
RH 95% or less