OM 10 micro-ohmmeter performs 4-wire measurements of inductive and non-inductive resistances up to 50 KΩ with continuous or pulse measuring current up to 100 mA. It works at a resolution of 10 µΩ and an accuracy of 0.1%. The accuracy of measurements is also ensured by the automatic compensation of thermal electromagnetic forces, the compensation of ambient temperature (internal sensor) at 20°C and by the selection of the metal type.
OM 10 can also measure DC and AC voltages up to 400 V.
OM 10 micro-ohmmeter performs 4-wire measurements of inductive and non-inductive resistances up to 50 KΩ with continuous or pulse measuring current up to 100 mA. It works at a resolution of 10 µΩ and an accuracy of 0.1%. The accuracy of measurements is also ensured by the automatic compensation of thermal electromagnetic forces, the compensation of ambient temperature (internal sensor) at 20°C and by the selection of the metal type.
OM 10 can also measure DC and AC voltages up to 400 V.
Key-features :
- 4-wire resistance measurement up to 50 K at 10µΩ resolution and 0.1% accuracy
- Voltage measurement up to 400 V at 1 V resolution and 1% accuracy
- Continuous measurement current for inductive loads: Coils, transformers, motor windings, twisted cables…
- Pulse measurement current for non-selfic resistances: Earth bonding, ground continuity, contact resistances, non-twisted cables…
- Automatic compensation of thermal E.M.F.
- Temperature compensation
- Metal temperature compensation
Applications :
- Cable resistance and resistivity measurement
- Inductive resistance measurement (transformers, motors…)
- Contact resistance measurement (connectors, switches, relays…)
- Metallisation, earth bounding and ground continuity tests
- Tests of low power electrical components (fuses…)
- Tests of heat-sensitive devices
- Tests of surface states, mechanical bonds and materials
Functionalities | OM 10 |
Measurement | Low resistance value measurement in 4-wire configuration Inductive and non-inductive resistances DC and AC voltage |
High accuracy | 0.05 % RDG |
Current | Pulse or DC current |
Range: | Up to 50 KΩ |
Resolution | 10 µΩ |
Protection | Up to 250 V / Up to 430 V for voltage measurement |
Power supply | Mains and internal battery |
Communication modes | RS 232 |
Resistance calculation | Automatic compensation for EMFs and ambient temperature Display of temperature compensated resistance value Choice of metal temperature coefficient Choice and display of ambient temperature (measured with integrated sensor) |
Additional functions | Calibration without internal adjustment |
Applications: | Cable resistance, motors, contact resistance, metallisation and earth bonding resistance, transformers… |