Features
|
|
|
Improved Production Efficiency Reduces Measurement and Inspection Time
The MS9740A is a benchtop optical spectrum analyzer covering 600 nm to 1750 nm. The MS9740A offers high-performance optical resolution and high-speed measurements to support the needs of device and component manufacturers.
Waveform data can be measured and transmitted to external control equipment within 0.2s — a 5x improvement over previous Anritsu models.
This enables optical device manufacturers to reduce costs and time for testing and analysis by:
– Improving sweep speed (measurement time), and
– Accelerating data transfer to control equipment after measurement
Combine these overall improvements with the ability to do basic analysis, and the result is excellent return on investment and improved productivity.
Specification | Details |
Applicable optical fiber | 10 µm/ 125 µm SM fiber (ITU-T G.652), 50 µm/ 125 µm GI fiber |
Optical connector | User replaceable: FC, SC, ST, DIN, LC |
Wavelength measurement range | 600 nm to 1750 nm |
Wavelength accuracy *1 | ±20 pm (1520 nm to 1620 nm, Resolution: 0.03 to 0.2 nm) *2, ±100 pm (1520 nm to 1620 nm, Resolution: 0.5, 1.0 nm) *2 |
Setting resolution | 0.03, 0.05, 0.07, 0.1, 0.2, 0.5, 1.0 nm (RBW: 3dB optical filter: transmission bandwidth) |
Maximum input power | +23 dBm (Optical ATT: On) |
Dynamic range *1 | High dynamic range: 60 dB (0.4 nm from peak wavelength) Normal dynamic range: 58 dB (0.4 nm from peak wavelength) [Resolution: 0.05 nm, 1550 nm, Optical Att: Off, 20 to 30°C] |
Sweep | Sweep width: 0.2 nm to 1200 nm, 0 nm Sweep speed: ≤0.2 sec. (span 5 nm) [Resolution: 0.1 nm, VBW=10 kHz, Normal dynamic range, center 1550 nm, sweep start to stop, no optical input, sampling point: ≤501] |
Dimensions and mass | 426 (W) x 177 (H) x 350 (D) mm (excluding projections) ≤15.0 kg (without option) |
*1: Using SM fiber (ITU-T G.652), after 2 hours of warm-up (the Repeat sweeping performed at Span 100 nm or more and VBW 10 kHz or more during the warm-up operation), after Auto Align, at stable room temperature.
*2: Built-in option 002, after Wl cal (ref) wavelength calibration execution, at stable room temperature.