ITI 17 is a programmable digital panel meter 96 x 48 mm for process and temperature by resistive probes and thermocouple measurements. The software enables additional linearization to be set up for thermocouples type B, C, E, L, Mo, N, PI, R, U and Pt100, Pt25 and Pt50. The connection of the probes is done using a removable screw connector.
- Panel meter size: 96 mm x 48 mm
- IP65 protected
- Two-color display for color change on alarm or pre-alarm
- Process signal measurement (0/4-20 mA, 0/100 mV /1 V / 10 V)
- Temperature measurement by Pt100 from -200°C to +850°C (0.01°C resolution up to 200°C) or by thermocouple type K, T, J or S
- Resistance measurement (10 mOhms to 400 Ohms)
- 2 different power supplies available:
- 90 to 260 V~ 47/420 Hz and 115 to 300 V-
- 20 to 60 V- and 18 to 44 V~ 47/420 Hz
3 versions:
- Version 0 : Panel meter only
- Version 1 : 2 limit relays, 1 programmable analogue output (4-20 mA or 0-10 V), 1 digital output RS232/485, MODBUS/ASCII
- Version 2 : 2 limit relays
Process signal measurement indicator
Type | Range | Actual range | Resolution | Accuracy (1 year) (23°C ± 5°C) |
Comments |
DC voltage | 60 mV 100 mV 1 V 10 V 100 V |
-6 to 75 mV -15 to +160 mV -0.1 to +1.2 V -1 to +12 V -10 to +120 V |
1 µV 10 μV 100 µV 1 mV 10 mV |
0.1% + 10 μV 0.1% + 20 μV 0.1% + 200 μV 0.1% + 2 mV 0.1% + 20 mV |
Input impedance 10 MΩ 10 MΩ 1.11 MΩ 10 MΩ |
DC current | 20 mA 4-20 mA |
-2 to 24 mA +3.2 to 24 mA |
1 µA | 0.1% + 4 µA | Loop power: 24 V ± 10 % @ 20 mA Applicable maximum current: 50 mA |
Resistance | 150 Ω 400 Ω |
0 to 160 Ω 0 to 420 Ω |
10 mΩ | 0.1% + 20 mΩ 0.1% + 50 mΩ |
3 wire measurement: add 0.05 % L/Ω of leads |
RTD temperature indicator
Range | Actual range | Resolution | Accuracy over 1 year (23°C ± 5°C) |
Comments |
150°C | -200 to +150°C | 0.01°C | 0.1% RDG + 0.05°C | 3-wire measurement: add (0.05% RDG + 0.15°C)/Ω of leads |
850°C | -200 to +850°C | 0.1°C | 0.1% RDG + 0.2°C |
Linearization compliant to EN 60751/1995 standard for platinum probe 100 Ω at 0°C and international scale of temperature ITS 90.
Thermocouple temperature indicator
Type | Range (1) | Resolution | Accuracy over 1 year (23°C ± 1°C) (2) |
Tc K | -200 to -100°C -100 to +1370°C |
0.1°C | ± 0.5% RDG 0.1% RDG + 0.3°C |
Tc T | -200 to -100°C -100 to +400°C |
0.1°C | ± 0.5% RDG 0.1% RDG + 0.3°C |
Tc J | -200 to -100°C -100 to +1200°C |
0.1°C | ± 0.4% RDG 0.1% RDG + 0.2°C |
Tc S | -50 to +300°C +300 to 1768°C |
1°C | ± 3°C 0.1% RDG + 1°C |
Uncertainty due to the internal reference junction: ± 1°C
Uncertainty due to the use of an external reference junction (AN8002): ± 0.2°C typical and 0.5°C max.
Linearization compliant to EN 60584-1/1995 standard and international scale of temperature ITS 90.
Possibility to add other thermocouples via software